Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
The primary spectrometer of the cold time-of-flight instrument IN5 at the ILL has been upgraded over the last 2 years. After commissioning in spring 2003 the absolute gain in flux when compared to the former spectrometer setting has been measured by gold-foil activation. Within standard configuration, namely, an incident wavelength of
λ=5
A
̊
and chopper speeds chosen such that they correspond to identical energy resolution
(δE≃100
μeV)
and time frame, the flux has been found an order of magnitude higher on the upgraded instrument. Apart from the gain in flux the new spectrometer allows to work with appreciably improved resolution (up to a factor 2) at identical wavelength.