Failure location predictoin by finite element analysis for an additive manufactured mandible implant
Ist Teil von
  • Medical engineering & physics, 2015, Vol.37 (9), p.862
Erscheinungsjahr
2015
Quelle
Alma/SFX Local Collection
Beschreibungen/Notizen
  • In order to reconstruct a patient with a bone defect in the mandible, a porous scaffold attached to a plate, both in a titanium alloy, was designed and manufactured using additive manufacturing. Regrettably, the implant fractured in vivo several months after surgery. The aim of this study was to investigate the failure of the implant and show a way of predicting the mechanical properties of the implant before surgery. All computed tomography data of the patient were preprocessed to remove metallic artefacts with metal deletion technique before mandible geometry reconstruction. The three-dimensional geometry of the patient's mandible was also reconstructed, and the implant was fixed to the bone model with screws in Mimics medical imaging software. A finite element model was established from the assembly of the mandible and the implant to study stresses developed during mastication. The stress distribution in the load-bearing plate was computed, and the location of main stress concentration in the plate was determined. Comparison between the fracture region and the location of the stress concentration shows that finite element analysis could serve as a tool for optimizing the design of mandible implants.
Sprache
Englisch
Identifikatoren
ISSN: 1873-4030, 1350-4533
DOI: 10.1016/j.medengphy.2015.06.001
Titel-ID: cdi_swepub_primary_oai_DiVA_org_uu_262683
Format

Weiterführende Literatur

Empfehlungen zum selben Thema automatisch vorgeschlagen von bX