Multicenter Trial of a Tubeless, On-Body Automated Insulin Delivery System With Customizable Glycemic Targets in Pediatric and Adult Participants With Type 1 Diabetes
Ist Teil von
Diabetes care, 2021-07, Vol.44 (7), p.1630-1640
Ort / Verlag
United States: American Diabetes Association
Erscheinungsjahr
2021
Quelle
EZB-FREE-00999 freely available EZB journals
Beschreibungen/Notizen
Advances in diabetes technology have transformed the treatment paradigm for type 1 diabetes, yet the burden of disease is significant. We report on a pivotal safety study of the first tubeless, on-body automated insulin delivery system with customizable glycemic targets.
This single-arm, multicenter, prospective study enrolled 112 children (age 6-13.9 years) and 129 adults (age 14-70 years). A 2-week standard therapy phase (usual insulin regimen) was followed by 3 months of automated insulin delivery. Primary safety outcomes were incidence of severe hypoglycemia and diabetic ketoacidosis. Primary effectiveness outcomes were change in HbA
and percent time in sensor glucose range 70-180 mg/dL ("time in range").
A total of 235 participants (98% of enrolled, including 111 children and 124 adults) completed the study. HbA
was significantly reduced in children by 0.71% (7.8 mmol/mol) (mean ± SD: 7.67 ± 0.95% to 6.99 ± 0.63% [60 ± 10.4 mmol/mol to 53 ± 6.9 mmol/mol],
< 0.0001) and in adults by 0.38% (4.2 mmol/mol) (7.16 ± 0.86% to 6.78 ± 0.68% [55 ± 9.4 mmol/mol to 51 ± 7.4 mmol/mol],
< 0.0001). Time in range was improved from standard therapy by 15.6 ± 11.5% or 3.7 h/day in children and 9.3 ± 11.8% or 2.2 h/day in adults (both
< 0.0001). This was accomplished with a reduction in time in hypoglycemia <70 mg/dL among adults (median [interquartile range]: 2.00% [0.63, 4.06] to 1.09% [0.46, 1.75],
< 0.0001), while this parameter remained the same in children. There were three severe hypoglycemia events not attributable to automated insulin delivery malfunction and one diabetic ketoacidosis event from an infusion site failure.
This tubeless automated insulin delivery system was safe and allowed participants to significantly improve HbA
levels and time in target glucose range with a very low occurrence of hypoglycemia.