Comparison of procedural efficacy and biophysical parameters between two competing cryoballoon technologies for pulmonary vein isolation: Insights from an initial multicenter experience
Ist Teil von
  • Journal of cardiovascular electrophysiology, 2021-03, Vol.32 (3), p.580-587
Ort / Verlag
United States: Wiley Subscription Services, Inc
Erscheinungsjahr
2021
Link zum Volltext
Quelle
Alma/SFX Local Collection
Beschreibungen/Notizen
  • Introduction Recently a novel cryoballoon system (POLARx, Boston Scientific) became available for the treatment of atrial fibrillation. This cryoballoon is comparable with Arctic Front Advance Pro (AFA‐Pro, Medtronic), however, it maintains a constant balloon pressure. We compared the procedural efficacy and biophysical characteristics of both systems. Methods One hundred and ten consecutive patients who underwent first‐time cryoballoon ablation (POLARx: n = 57; AFA‐Pro: n = 53) were included in this prospective cohort study. Results Acute isolation was achieved in 99.8% of all pulmonary veins (POLARx: 99.5% vs. AFA‐Pro: 100%, p = 1.00). Total procedure time (81 vs. 67 min, p < .001) and balloon in body time (51 vs. 35 min, p < .001) were longer with POLARx. After a learning curve, these times were similar. Cryoablation with POLARx was associated with shorter time to balloon temperature −30°C (27 vs. 31 s, p < .001) and −40°C (32 vs. 54 s, p < .001), lower balloon nadir temperature (−55°C vs. −47°C, p < .001), and longer thawing time till 0°C (16 vs. 9 s, p < .001). There were no differences in time‐to‐isolation (TTI; POLARx: 45 s vs. AFA‐Pro 43 s, p = .441), however, POLARx was associated with a lower balloon temperature at TTI (−46°C vs. −37°C, p < .001). Factors associated with acute isolation differed between groups. The incidence of phrenic nerve palsy was comparable (POLARx: 3.5% vs. AFA‐Pro: 3.7%). Conclusion The novel cryoballoon is comparable to AFA‐Pro and requires only a short learning curve to get used to the slightly different handling. It was associated with faster cooling rates and lower balloon temperatures but TTI was similar to AFA‐Pro.
Sprache
Englisch
Identifikatoren
ISSN: 1045-3873
eISSN: 1540-8167
DOI: 10.1111/jce.14915
Titel-ID: cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_7986676

Weiterführende Literatur

Empfehlungen zum selben Thema automatisch vorgeschlagen von bX