Design and application of adjustable triangle external fixator for the treatment of calcaneal fractures
Ist Teil von
Zhongguo gu shang, 2009-12, Vol.22 (12), p.892
Ort / Verlag
China
Erscheinungsjahr
2009
Link zum Volltext
Quelle
MEDLINE
Beschreibungen/Notizen
To design a kind of adjustable triangle external fixator, and to use it for treating calcaneal fractures.
The external fixator was installed into extremity near the both sides of three Kirschner wires crossing the calcaneal tubercle, the lower end of the tibia and the first cuneiform. The support rod length was adjusted to ensure the Kirschner wires affecting the calcaneal tubercle, so the Böhler angle and anteroposterior diameter of the calcaneus were recovered. From April in 2003 to April in 2008, the external fixators were used to treat 34 patients with calcaneal fractures, including 30 males and 4 females, ranging in age from 19 to 54 years, with an average of 36.7 years. The Böhler angles were -5 degree to 15 degree, with an average of 13.5 degrees. According to Sander's classification, 2 patients were type II, 20 patients were type III, and 12 patients were type IV.
Thirty-four patients were followed up, and the duration ranged from 3 months to 4 years, averaged 25 months. The Böhler angle recovered to