A search for optimal x-ray spectra in iodine contrast media mammography
Ist Teil von
Physics in medicine & biology, 2005-07, Vol.50 (13), p.3143-3152
Ort / Verlag
England: IOP Publishing
Erscheinungsjahr
2005
Quelle
MEDLINE
Beschreibungen/Notizen
The aim of this work was to search for the optimal x-ray tube voltage and anode-filter combination in digital iodine contrast media mammography. In the optimization, two entities were of interest: the average glandular dose, AGD, and the signal-to-noise ratio, SNR, for detection of diluted iodine contrast medium. The optimum is defined as the technique maximizing the figure of merit, SNR2/AGD. A Monte Carlo computer program was used which simulates the transport of photons from the x-ray tube through the compression plate, breast, breast support plate, anti-scatter grid and image detector. It computes the AGD and the SNR of an iodine detail inside the compressed breast. The breast thickness was varied between 2 and 8 cm with 10-90% glandularity. The tube voltage was varied between 20 and 55 kV for each anode material (Rh, Mo and W) in combination with either 25 microm Rh or 0.05-0.5 mm Cu added filtration. The x-ray spectra were calculated with MCNP4C (Monte Carlo N-Particle Transport Code System, version 4C). A CsI scintillator was used as the image detector. The results for Rh/0.3 mmCu, Mo/0.3 mmCu and W/0.3 mmCu were similar. For all breast thicknesses, a maximum in the figure of merit was found at approximately 45 kV for the Rh/Cu, Mo/Cu and W/Cu combinations. The corresponding results for the Rh/Rh combination gave a figure of merit that was typically lower and more slowly varying with tube voltage. For a 4 cm breast at 45 kV, the SNR2/AGD was 3.5 times higher for the Rh/0.3 mmCu combination compared with the Rh/Rh combination. The difference is even larger for thicker breasts. The SNR2/AGD increases slowly with increasing Cu-filter thickness. We conclude that tube voltages between 41 and 55 kV and added Cu-filtration will result in significant dose advantage in digital iodine contrast media mammography compared to using the Rh/Rh anode/filter combination at 25-32 kV.