Dynamic response of diamond sensors to ionizing radiation beams
Ist Teil von
Sensors and actuators. A. Physical., 2011-11, Vol.171 (1), p.43-47
Ort / Verlag
Elsevier B.V
Erscheinungsjahr
2011
Quelle
ScienceDirect
Beschreibungen/Notizen
Prototypes of radiation detectors for clinical dosimetry were constructed using commercially available synthetic diamond. Polished diamond plates of detector grade and standard quality were used. The metallization scheme was a typical Ag–Ag contact for the detector grade diamond and a new metallization technique for the standard quality diamond. This was for determining which detector is more appropriate for radiotherapy dosimetry needs. The detector signal was analyzed as a function of bias voltage in the 0–500
V range both in absence of radiation and in a Co-60 gamma-ray beam. The response linearity was assessed in Co-60 gamma-ray beam for dose-rates in the range from 0.2 to 1.2
Gy
min
−1. Pre-irradiation effects were also investigated. Signal rise time, signal stability and reproducibility are reported and discussed.