A Compton imager for in-vivo dosimetry of proton beams—A design study
Ist Teil von
Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 2011, Vol.626, p.114-119
Ort / Verlag
Elsevier B.V
Erscheinungsjahr
2011
Quelle
Elsevier ScienceDirect Journals
Beschreibungen/Notizen
In-beam SPECT during therapeutic proton beam irradiation is a novel method for three dimensional in-vivo dose verification. For this purpose a Compton camera design is evaluated with respect to the special requirements and conditions that arise from this application. Different concepts are studied by means of simulation concerning the angular resolution and efficiency. It was found that a cadmium zinc telluride system can perform sufficiently well. For further evaluation the construction of a semiconductor scintillator hybrid system is under way.