Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
Ergebnis 11 von 29
Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 2011-06, Vol.642 (1), p.75-77
2011
Volltextzugriff (PDF)

Details

Autor(en) / Beteiligte
Titel
Improved scintillation proportionality and energy resolution of LaBr sub(3):Ce at 80 K
Ist Teil von
  • Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 2011-06, Vol.642 (1), p.75-77
Erscheinungsjahr
2011
Quelle
Elsevier ScienceDirect Journals
Beschreibungen/Notizen
  • Using highly monochromatic synchrotron X-rays in the energy range from 10.5 to 100 keV, the temperature dependence of the photon nonproportional response and energy resolution of LaBr sub(3) scintillators doped with 5% Ce super(3+) were studied at 80, 295, and 450 K. Improvement of the proportionality and better energy resolution was observed on lowering the temperature. This effect suggests that the already outstanding energy resolution of LaBr sub(3):Ce can be improved even further. It may also provide new clues to better understanding of the processes that cause nonproportionality of inorganic scintillator response. Temperature dependence of nonproportionality of LaBr sub(3):Ce was discovered Improvement of energy resolution was observed on lowering the temperature Outstanding energy resolution of LaBr sub(3):Ce scintillator can be improved even further Parameters responsible for nonproportionality of LaBr sub(3):Ce depend on temperature Our results demonstrate that nonproportionality is a property that can be improved.
Sprache
Englisch
Identifikatoren
ISSN: 0168-9002
DOI: 10.1016/j.nima.2011.04.006
Titel-ID: cdi_proquest_miscellaneous_889387998

Weiterführende Literatur

Empfehlungen zum selben Thema automatisch vorgeschlagen von bX