Lymphocyte subset analysis on frozen whole blood
Cytometry (New York, N.Y.), 1997-12, Vol.29 (4), p.340-350
Fiebig, Eberhard W.
Johnson, Delene K.
Hirschkorn, Dale F.
Knape, Charlene C.
Webster, H. Kyle
Lowder, James
Busch, Michael P.
1997
Volltextzugriff (PDF)
Details
Autor(en) / Beteiligte
Fiebig, Eberhard W.
Johnson, Delene K.
Hirschkorn, Dale F.
Knape, Charlene C.
Webster, H. Kyle
Lowder, James
Busch, Michael P.
Titel
Lymphocyte subset analysis on frozen whole blood
Ist Teil von
Cytometry (New York, N.Y.), 1997-12, Vol.29 (4), p.340-350
Ort / Verlag
Hoboken: Wiley Subscription Services, Inc., A Wiley Company
Erscheinungsjahr
1997
Quelle
MEDLINE
Beschreibungen/Notizen
An approach to perform lymphocyte subset analysis on frozen‐thawed whole blood (F/T WB) is described. WB from 24 human immunodeficiency virus type 1 (HIV‐1) seropositive individuals and 21 controls was analyzed fresh and after frozen storage (with or without dimethyl sulfoxide) at −80°C, in liquid nitrogen (LN2), and at −20°C. Analysis of F/T WB utilized 3‐color flow cytometry with CD45 and right angle light scatter gating. Absolute cell counts were obtained for 30 samples by using staining tubes containing internal bead standards [TruCount, Becton Dickinson Immunocytometry Systems (BDIS), San Jose, CA]. The mean difference between CD3+4+ percentages for F/T (−80°C storage for up to 1 year) and fresh WB was less than −0.2% (95% limits ±3%, P = 0.5) with 39 of 45 (87%) results falling within 2% of the fresh values (P = 0.74). Absolute CD3+4+ cell counts for F/T WB were generally lower than corresponding results for fresh aliquots (median difference was 33 cells/μl, P < 0.0001), but the results were highly correlated (r2 = 0.975, P < 0.0001). Results were more variable, although still highly correlated, for CD3+8+ cells, and with other freezing and storage conditions. It is concluded that lymphocyte subset analysis using F/T WB yields comparable results to fresh samples, which should prove useful for a number of practical applications. Cytometry 29:340–350, 1997. © 1997 Wiley‐Liss, Inc.
Sprache
Englisch
Identifikatoren
ISSN: 0196-4763
eISSN: 1097-0320
DOI: 10.1002/(SICI)1097-0320(19971201)29:4<340::AID-CYTO11>3.0.CO;2-U
Titel-ID: cdi_proquest_miscellaneous_79495337
Format
–
Schlagworte
AIDS/HIV
,
Blood Preservation
,
Blood Specimen Collection
,
Cell Survival
,
Cryopreservation
,
flow cytometry
,
Flow Cytometry - methods
,
Humans
,
Lymphocyte Count
,
lymphocyte subsets
,
Lymphocyte Subsets - cytology
,
Lymphocytes
,
TruCount method
,
whole blood