Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
An improved method for clenbuterol screening using high resolution selected ion recording
Ist Teil von
Rapid communications in mass spectrometry, 1995, Vol.9 (14), p.1395-1399
Ort / Verlag
London: Heyden & Son Limited
Erscheinungsjahr
1995
Quelle
Wiley Online Library All Journals
Beschreibungen/Notizen
A method has been developed using reliable GC derivatization techniques interfaced with a high resolution mass spectrometer. The method has proved successful for the detection of low levels (less than 1 ppm) of clenbuterol in complex biological matrices. Selected ion recording of two characteristic isotopic fragment ions provides a specific mode of detection by verifying the GC retention time of these ions and also by comparing their relative abundance. Analysis of urine samples demands higher mass spectrometric resolution, and 40000 (10% valley) was found to be a prerequisite for accurate integration of the drug‐related chromatographic peaks. The method developed is suitable for adaptation to a completely unattended automated routine incorporating sample injection, storage and retrieval of source tuning parameters, and data processing.