Response of Breasts to Different Stimulation Patterns of an Electric Breast Pump
Ist Teil von
Journal of human lactation, 2003-05, Vol.19 (2), p.179-186
Ort / Verlag
United States: SAGE Publications
Erscheinungsjahr
2003
Quelle
MEDLINE
Beschreibungen/Notizen
To test the effect on milk ejection, an electric breast pump was programmed to provide pumping patterns with frequencies of 45 to 125 cycles/min and vacuums of -45 to -273 mm Hg. The time taken for milk ejection to occur (measured using ultrasound to detect a dilation of a lactiferous duct in the opposite breast) in response to the current Medela electric breast pump pattern (45 cycles/min) was 147 ± 13 s. For patterns that more closely resemble the sucking frequency of an infant when it first attaches to the breast, milk ejection occurred between 136 ± 12 and 104 ± 10 s, although this difference was not statistically significant. Milk ejection in response to breastfeeding occurred after 56 ± 4 s. The applied vacuum affected the amount of milk that was removed up to 50 to 70 s after milk ejection but not the time for milk ejection. J Hum Lact. 19(2):179- 186.