Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
Acoustic power calibration of high-intensity focused ultrasound transducers using a radiation force technique
Ist Teil von
The Journal of the Acoustical Society of America, 2007-03, Vol.121 (3), p.1434-1439
Ort / Verlag
Woodbury, NY: Acoustical Society of America
Erscheinungsjahr
2007
Quelle
American Institute of Physics (AIP) Journals
Beschreibungen/Notizen
To address the challenges associated with measuring the ultrasonic power from high-intensity focused ultrasound transducers via radiation force, a technique based on pulsed measurements was developed and analyzed. Two focused ultrasound transducers were characterized in terms of an effective duty factor, which was then used to calculate the power during the pulse at high applied power levels. Two absorbing target designs were used, and both gave comparable results and displayed no damage and minimal temperature rise if placed near the transducer and away from the focus. The method yielded reproducible results up to the maximum pulse power generated of approximately
230
W
, thus allowing the radiated power to be calibrated in terms of the peak-to-peak voltage applied to the transducer.