Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
Ergebnis 11 von 13224
Journal of microscopy (Oxford), 2008-01, Vol.229 (1), p.115-126
2008

Details

Autor(en) / Beteiligte
Titel
New controlled environment vitrification system for preparing wet samples for cryo‐SEM
Ist Teil von
  • Journal of microscopy (Oxford), 2008-01, Vol.229 (1), p.115-126
Ort / Verlag
Oxford, UK: Blackwell Publishing Ltd
Erscheinungsjahr
2008
Link zum Volltext
Quelle
Wiley Online Library Journals Frontfile Complete
Beschreibungen/Notizen
  • Summary A new controlled environment vitrification system (CEVS) has been designed and constructed to facilitate examination by cryogenic scanning electron microscopy (Cryo‐SEM) of initial suspension state and of microstructure development in latex, latex–composite and other coatings while they still contain solvent. The new system has a main chamber with provisions for coating as well as drying, and for well‐controlled plunging into cryogen. An added subsidiary chamber holds samples for drying or annealing over minutes to days before they are returned to the main chamber and plunged from it. In the main chamber, samples are blade‐coated on 5 × 7 mm pieces of silicon wafer and held at selected temperature and humidity for successively longer times, either there or after transfer along a rail into the subsidiary chamber. They are then placed in the sample holder mounted on the plunge rod, so as to permit adjustment of the sample's attitude when it plunges, at controlled speed, into liquid ethane at its freezing point, to a chosen depth, in order to solidify the sample without significant shear or freezing artifacts. The entries of plunging samples and related sample holders into liquid ethane were recorded with a high‐speed, high‐resolution Photron digital camera. The data were interpreted with a new hypothesis about the width of the band of extremely rapid cooling by deeply subcooled nucleate boiling below the line of entry. Complementary cryo‐SEM images revealed that the freezing rate and surface shearing of a sample need to be balanced by adjusting the plunging attitude.

Weiterführende Literatur

Empfehlungen zum selben Thema automatisch vorgeschlagen von bX