Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
Ergebnis 10 von 80

Details

Autor(en) / Beteiligte
Titel
Dual-chamber defibrillators reduce clinically significant adverse events compared with single-chamber devices: results from the DATAS (Dual chamber and Atrial Tachyarrhythmias Adverse events Study) trial
Ist Teil von
  • Europace (London, England), 2008-05, Vol.10 (5), p.528-535
Ort / Verlag
England: Oxford University Press
Erscheinungsjahr
2008
Quelle
Oxford Journals 2020 Medicine
Beschreibungen/Notizen
  • Aims This randomized trial evaluated clinically significant adverse events (CSAEs), in patients implanted with dual-chamber (DC) vs. single-chamber (SC) implantable cardioverter defibrillator (ICD). DC-ICD had atrial tachyarrhythmia (AT) therapy capabilities. Strict programming recommendations were reinforced. Methods and results Patients with conventional SC-ICD indication were randomized to DC-ICD, SC-ICD, or a DC-ICD programmed as an SC-ICD (SC-simulated) and followed for 16 months. Patients in the DC and SC-simulated groups crossed over after 8 months. The primary endpoint was a composite of CSAE: all-cause mortality; invasive intervention; hospitalization (>24 h) for cardiovascular causes; inappropriate shocks (two or more episodes); and sustained symptomatic AT lasting >48 h. The outcome variable was a pre-specified score that corrected for clinical severity and follow-up duration. Three hundred and thirty-four patients were analysed (DC-ICD, n = 112; SC-ICD, n = 111; SC-simulated, n = 111). The mean left ventricular ejection fraction was 0.36 ± 0.13, 69% were in functional class ≥II. CSAE occurred in 65 DC-ICD, 82 SC-ICD, and 84 SC-simulated patients. The outcome variable was 33% lower in the DC-ICD group (OR 0.31; 95% CI 0.14-0.67; P = 0.0028). Mortality was 4% in DC, 9% in SC, and 10% in SC-simulated. Conclusion In patients with a standard SC-ICD indication, DC-ICD was associated with less CSAE when compared with SC-ICD.

Weiterführende Literatur

Empfehlungen zum selben Thema automatisch vorgeschlagen von bX