This study presents minimum 6-year follow-up data on the survival and satisfaction of an uncemented modular revision femoral system, following on from our previously published earlier results.
We retrospectively reviewed all revision hip arthroplasties performed at our institution between January 2005 and October 2012, using a single modular femoral revision system. Patient-reported outcomes were collected (satisfaction score and Oxford Hip Score). Preoperative and postoperative radiographs were reviewed for stem subsidence, and Kaplan–Meier analysis was performed for survival. A total of 115 femoral revisions were performed in 106 patients.
All-cause survival was 82% (95% confidence interval 74 to 91%) at 10.8 years, and 96% (95% confidence interval 90 to 100%) excluding septic failure. Of the 19 cases requiring reoperation, 16 were for infection, 2 for aseptic loosening, and 1 for mechanical failure. At final follow-up, 88.5% of patients were “satisfied” or “very satisfied”.
This study showed excellent clinical results of a commonly used revision hip stem with at least 10 years follow-up. Satisfaction rates were high, with few aseptic failures. Stem subsidence was more common in revisions for infection, but did not correlate with lower satisfaction scores.