Comparison of the accuracy of dental implant placement using static and dynamic computer-assisted systems: an in vitro study
Ist Teil von
Journal of stomatology, oral and maxillofacial surgery, 2021-09, Vol.122 (4), p.343-348
Ort / Verlag
France: Elsevier Masson SAS
Erscheinungsjahr
2021
Link zum Volltext
Quelle
Alma/SFX Local Collection
Beschreibungen/Notizen
The study aimed to compare the accuracy of implant placement between static and dynamic computer-assisted systems (CAS) in a partially edentulous mandible model.
A total of 80 implants was placed in mandible models. The implants were placed using either static or dynamic computer-assisted system. Deviations of implant hex, apex and angulation were measured between preoperative planning and postoperative CBCT in planning software.
The mean deviations of implant hex, apex and angulation in static CAS group were 1.15 ± 0.34 mm, 1.37 ± 0.38 mm and 2.60 ± 1.11 degree, while in dynamic CAS group were 0.40 ± 0.41 mm, 0.34 ± 0.33 mm and 0.97 ± 1.21 degree, respectively. Implant placement with dynamic CAS showed less deviations of shoulder, apex and angulation than with static CAS significantly.
The implant accuracy using CAS system could be influenced by the guiding technique.