Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
Ergebnis 3 von 29

Details

Autor(en) / Beteiligte
Titel
High-Frequency Irreversible Electroporation Using 5,000-V Waveforms to Create Reproducible 2- and 4-cm Ablation Zones—A Laboratory Investigation Using Mechanically Perfused Liver
Ist Teil von
  • Journal of vascular and interventional radiology, 2020-01, Vol.31 (1), p.162-168.e7
Ort / Verlag
United States: Elsevier Inc
Erscheinungsjahr
2020
Beschreibungen/Notizen
  • To investigate if high-frequency irreversible electroporation (H-FIRE) treatments can be delivered at higher voltages and with greater energy delivery rates than currently implemented in clinical irreversible electroporation protocols. Treatments using 3,000 V and 5,000 V were administered to mechanically perfused ex vivo porcine liver via a single applicator and grounding pad (A+GP) as well as a 4-applicator array (4AA). Integrated energized times (IET) 0.01–0.08 seconds and energy delivery rates 25–300 μs/s were investigated. Organs were preserved at 4°C for 10–15 hours before sectioning and gross analysis using a metabolic stain to identify the size and shape of ablation zones. A+GP ablations measured between 1.6 cm and 2.2 cm, which did not increase when IET was increased from 0.02 seconds to 0.08 seconds (P > .055; range, 1.9–2.1 cm). Changes in tissue color and texture consistent with thermal damage were observed for treatments with energy delivery rates 50–300 μs/s, but not for treatments delivered at 25 μs/s. Use of the 4AA with a 3-cm applicator spacing resulted in ablations measuring 4.4–4.9 cm with energy delivery times of 7–80 minutes. H-FIRE treatments can rapidly and reproducibly create 2-cm ablations using an A+GP configuration. Treatments without thermal injury were produced at the expense of extended treatment times. More rapid treatments resulted in ablations with varying degrees of thermal injury within the H-FIRE ablation zone. Production of 4-cm ablations is possible using a 4AA.
Sprache
Englisch
Identifikatoren
ISSN: 1051-0443
eISSN: 1535-7732
DOI: 10.1016/j.jvir.2019.05.009
Titel-ID: cdi_proquest_miscellaneous_2293012688
Format

Weiterführende Literatur

Empfehlungen zum selben Thema automatisch vorgeschlagen von bX