Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
Ergebnis 8 von 204

Details

Autor(en) / Beteiligte
Titel
Japan ablation registry: cryoablation in atrioventricular nodal reentrant tachycardia (“JARCANRET study”): results from large multicenter retrospective investigation
Ist Teil von
  • Journal of interventional cardiac electrophysiology, 2020-09, Vol.58 (3), p.289-297
Ort / Verlag
New York: Springer US
Erscheinungsjahr
2020
Link zum Volltext
Quelle
SpringerLink (Online service)
Beschreibungen/Notizen
  • Purpose Cryoenergy has been demonstrated to be a safe alternative to radiofrequency ablation for catheter ablation of atrioventricular nodal reentrant tachycardia (AVNRT). This study aimed to evaluate the safety and efficacy of cryoablation in patients with AVNRT. Methods A multicenter retrospective study was performed. Two hundred eighty-three consecutive patients with AVNRT underwent cryoablation. Cryomapping at − 30 °C and − 80 °C was performed to predict cryoablation outcome and ascertain antegrade conduction. Cryoenergy was delivered subsequently at the same spot (cryoablation at − 80 °C) for 240 s. Results Ablation procedure was acutely successful in 281 out of 283 patients (99.3%). Of note, 22 patients (10.1%) had transient AV block during the cryoablation, but no injurious effects on AV conduction were provoked during cryomapping. During a follow-up period of 367 ± 35 days, the recurrence rate was 3.9% (11 out of 281). There were no significant differences among the patients with a complete elimination of slow pathway conduction, AH jump without an echo beat, and AH jump with a single echo beat, in terms of the long-term recurrence of AVNRT. Conclusions Cryoablation of AVNRT appears to be effective both acutely and during the long-term with a minimal risk of unwanted injuries to the conduction system. It seems to be important to monitor the antegrade conduction during cryoenergy applications, even when cryomapping demonstrates a safe location for cryoablation. The recurrence rate of AVNRT did not differ according to the properties of the residual slow pathway conduction.
Sprache
Englisch
Identifikatoren
ISSN: 1383-875X
eISSN: 1572-8595
DOI: 10.1007/s10840-019-00585-0
Titel-ID: cdi_proquest_miscellaneous_2267749710

Weiterführende Literatur

Empfehlungen zum selben Thema automatisch vorgeschlagen von bX