Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
Specimen‐thickness effects on transmission Kikuchi patterns in the scanning electron microscope
Ist Teil von
Journal of microscopy (Oxford), 2014-06, Vol.254 (3), p.129-136
Ort / Verlag
England: Wiley Subscription Services, Inc
Erscheinungsjahr
2014
Quelle
Wiley Online Library
Beschreibungen/Notizen
Summary
We report the effects of varying specimen thickness on the generation of transmission Kikuchi patterns in the scanning electron microscope. Diffraction patterns sufficient for automated indexing were observed from films spanning nearly three orders of magnitude in thickness in several materials, from 5 nm of hafnium dioxide to 3 μm of aluminum, corresponding to a mass‐thickness range of ∼5 to 810 μg cm–2. The scattering events that are most likely to be detected in transmission are shown to be very near the exit surface of the films. The energies, spatial distribution and trajectories of the electrons that are transmitted through the film and are collected by the detector are predicted using Monte Carlo simulations.