Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
To establish an automatic and online microfluidic chip-mass spectrometry (chip-MS) system, a device was designed and fabricated for microsampling by a hybrid capillary. The movement of the capillary was programmed by a computer to aspirate samples from different microfluidic channels in the form of microdroplets (typically tens of nanoliters in volume), which were separated by air plugs. The droplets were then directly analyzed by MS via paper spray ionization without any pretreatment. The feasibility and performance were demonstrated by a concentration gradient experiment. Furthermore, after eliminating the effect of nonuniform response factors by an internal standard method, determination of the association constant within a noncovalent protein-protein complex was successfully accomplished with the MS-based titration indicating the versatility and the potential of this novel platform for widespread applications.