Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
Conceptual design of a proton computed tomography system for applications in proton radiation therapy
Ist Teil von
IEEE transactions on nuclear science, 2004-06, Vol.51 (3), p.866-872
Ort / Verlag
New York: IEEE
Erscheinungsjahr
2004
Quelle
IEEE
Beschreibungen/Notizen
Proton computed tomography (pCT) has the potential to improve the accuracy of dose calculations for proton treatment planning, and will also be useful for pretreatment verification of patient positioning relative to the proton beam. A design study was performed to define the optimal approach to a pCT system based on specifications for applications in proton therapy. Conceptual and detailed design of a pCT system is presented; the system consists of a silicon-based particle tracking system and a crystal calorimeter to measure energy loss of individual protons. We discuss the formation of pCT images based on the reconstruction of volume electron density maps and the suitability of analytic and statistical algorithms for image reconstruction.