Characterization and performance of co-axial HPGe detectors
Ist Teil von
Journal of radioanalytical and nuclear chemistry, 2024, Vol.333 (6), p.3123-3135
Ort / Verlag
Cham: Springer International Publishing
Erscheinungsjahr
2024
Quelle
Alma/SFX Local Collection
Beschreibungen/Notizen
High purity germanium (HPGe) detectors are deployed globally for gamma-radiation spectroscopy due to their superior energy resolution. In this work, the essential characteristics of n and p-type HPGe detectors, such as energy resolution, efficiency, minimum detectable activity (MDA), and peak shape were studied for the purpose of characterization and performance optimization. The results are obtained for various source-detector configurations in a wide energy range of 40–1408 keV using gamma sources, such as
109
Cd,
57
Co,
137
Cs,
54
Mn,
65
Zn,
60
Co, and
152
Eu. Scanning (distance, lateral, and radial) of the detectors was performed using different gamma sources to understand the orientation of the crystal with its active volume and counting efficiency and to characterize the geometry in detail. The ambient background around the n-type HPGe was reduced using Pb-shielding. As a result, an 85.85% suppression was observed in the mean integral window of 40–2700 keV. The characterization and performance tests of the detectors convincingly suggest that both the detectors can be deployed for environmental radioactivity explorations.