Reprint of “X-ray spectrometry applied for determination of linear attenuation coefficient of tissue-equivalent materials”
Ist Teil von
Radiation physics and chemistry (Oxford, England : 1993), 2020-02, Vol.167, p.108553, Article 108553
Ort / Verlag
Oxford: Elsevier Ltd
Erscheinungsjahr
2020
Quelle
Alma/SFX Local Collection
Beschreibungen/Notizen
Resin-based materials equivalent to water were developed and four different samples were obtained. The linear attenuation coefficients of all samples was evaluated using X-ray spectrometry with primary and transmitted beams using voltages at the ranging from 60 to 120 kV. The experimental measured values were compared with theoretical reference values to water and with that obtained using the Least Square Method algorithm methodology (method applied to diagnostic radiology and radiotherapy). Our results show that differences between the measured values and the target μ(E) was lower than 7% ± 0.3 in the energy range from 20 to 80 keV. These results enable to consider that the material developed and produced is a good option to be used as a water-equivalent material and the experimental method adequate to its quantitative evaluation.
•Resin-based materials equivalent to water were developed and produced.•The developed and produced material is a good option to be used as a water-equivalent material.•These results enable to consider experimental method adequate to its quantitative evaluation.•The results allow the evaluation of the sample which better represents the reference material.