Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
Central foveal thickness (CFT) measurements from optical coherence tomography (OCT) scans provide a precise measure of severity of pathologic changes in the fovea, progress of disease and response to treatment. Although these measures are additionally valuable to assess foveal development in infants, their reproducibility is not known. The goal of this retrospective study is to evaluate the variation and reproducibility of CFT measurements using handheld spectral-domain OCT (hh-SDOCT) in supine infants compared to conventional adult tabletop imaging.
Imaging sessions with multiple macular, volume scans in one eye were selected for analysis from two participant groups: Group 1, 25 imaging sessions from 21 preterm infants without macular edema imaged supine in the nursery using hh-SDOCT (Leica/Bioptigen Envisu C2300, RTP, NC); Group 2, 25 imaging sessions from 25 adults imaged using tabletop Bioptigen SDOCT. For each imaging session, three macular OCT volumes with acceptable image quality were selected for analysis. CFTs were measured using a customized script for automatic segmentation. An expert grader and a typical grader corrected the segmentation lines for the central foveal frame. Coefficient of variations (CV) and intraclass correlation coefficients (ICC) were calculated for graders and systems and compared to the previous literature on OCT reproducibility.
CFT measurements were repeatable and reproducible for both handheld and tabletop SDOCT systems. For handheld, grader ICC (CI) and mean CV were 0.94 (0.90-0.97) and 3.8 (typical) and 0.98 (0.96-0.99) and 2.9 (expert), and for tabletop were 0.91(0.83-0.96) and 2.1 (typical) and 0.92 (0.86-0.96) and 1.9 (expert). Intergrader reproducibility of handheld and tabletop SDOCT systems were ICC(CI) 0.97 (0.95-0.98) and 0.93 (0.89-0.96) respectively, and both are comparable to previously reported reproducibility of tabletop systems.
Handheld SDOCT is a reproducible instrument to measure foveal thicknesses in supine infants. It can be used in clinical research to evaluate foveal changes during retinal development and pathological conditions.