Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
United States: Materials Research Society, Warrendale, PA (United States)
Erscheinungsjahr
1998
Beschreibungen/Notizen
Continued improvements in the manufacturing of Cd{sub 1{minus}x}Zn{sub x}Te (CZT) material have resulted in a practical thermoelectrically cooled X-ray and gamma-ray detector of very high energy resolution. A high resolution spectroscopy system was used to measure the Fano factor in CZT at temperatures down to {minus}40 C. The best resolution of the 5.9 keV {sup 55}Fe peak was measured to be 188 eV FWHM, while the best resolution of the 59.5 keV {sup 241}Am peak was measured to be 482 eV FWHM. The minimum measured Fano factor was 0.082, with several measurements yielding a value of 0.089 {+-} 0.005. With a resolution of 4.2 keV FWHM for the 662 keV peak of {sup 137}Cs, these detectors demonstrate excellent performance in detecting X-rays and gamma rays.