A field-programmable lab-on-a-chip with built-in self-test circuit and low-power sensor-fusion solution in 0.35μm standard CMOS process
Ist Teil von
2015 IEEE Asian Solid-State Circuits Conference (A-SSCC), 2015, p.1-4
Ort / Verlag
IEEE
Erscheinungsjahr
2015
Quelle
IEEE Xplore
Beschreibungen/Notizen
This paper presents a field programmable lab-on-a-chip (FPLOC) with microfluidics actuation, droplet location/ category readback, built-in-self-test (BIST), and bioassays detection. This FPLOC is composed of 1,800 MEDA microelectrodes and integrates a set of microfluidic operations to perform droplet moving/cutting/mixing. Besides, capacitive sensing circuits are considered for system automation, droplets classification, malfunctioning microelectrodes and surface flatness testing, and biomedical detection. Implemented in 0.35μm standard CMOS process in 7.4mm2, the FPLOC demonstrates microfluidic operations under 25V at 1KHz, 1.3fF resolution for droplet location/category/chip BIST maps readback at 1MHz, and 0.34fF sensitivity of capacitive measurement window for biomedical detection at 1KHz.