Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
Ergebnis 26 von 276

Details

Autor(en) / Beteiligte
Titel
Tribocorrosion of Polyethylene/Cobalt Contact Combined with Real-Time Fluorescence Assays on Living Macrophages: Development of A Multidisciplinary Biotribocorrosion Device
Ist Teil von
  • Biotribology (Oxford), 2019-06, Vol.18 (19), p.100091, Article 100091
Ort / Verlag
Elsevier Ltd
Erscheinungsjahr
2019
Quelle
Alma/SFX Local Collection
Beschreibungen/Notizen
  • The test conditions currently used in biotribocorrosion devices often differ greatly from the physiological conditions of joint replacements, contributing to discrepancies between the simulated and actual life span of joint replacements. In this study, a multidisciplinary biotribocorrosion device was developed based on the limitations of existing tribocorrosimeters. The set-up enables corrosion measurements to be simultaneously performed with real-time visualization of living cells using fluorescence microscopy under dynamic loads and movements. The device was configured to simulate the joint contact of ankle prostheses, and the wear of ultra-high-molecular-weight polyethylene/cobalt alloy (CoCrMo) implants surrounded by murine macrophages was tested. Various characterization techniques (non-contact optical profilometry, scanning and fluorescence electron microscopy and quantitative analyses of metal ions and pro-inflammatory cytokines) were combined in-depth multidisciplinary study. Two experimental conditions were used to promote the production of either polyethylene wear particles or metal ions. The first results indicated two distinct tribocorrosion mechanisms: 1) adhesive wear coupled with slow ionic depassivation of the cobalt alloy. The main degradation products were micrometric spherical polyethylene particles that seem to have little impact effect on the metabolic activity of the macrophages. 2) Ionic wear with the production of small, fibrillar polyethylene particles was observed. The production of metal ions, mainly chromium, was the predominant degradation process. The cytotoxicity of the chromium ions was evaluated based on the secretion of pro-inflammatory cytokines (prostaglandin E2). Our findings indicate that simulated conditions that result in low mechanical wear but high ions release appear to be more harmful to cells. [Display omitted] •A tribometer was adapted into a multidisciplinary biotribocorrosion device.•The simultaneous use of tribocorrosion and bioassays discriminates the synergistic mechanism.•Polyethylene/cobalt alloy contact is shown to induce adhesive or ionic wear.•Metal ions decreases the proliferation and adherence of living macrophages.
Sprache
Englisch
Identifikatoren
ISSN: 2352-5738
eISSN: 2352-5738
DOI: 10.1016/j.biotri.2019.100091
Titel-ID: cdi_hal_primary_oai_HAL_hal_02133472v1

Weiterführende Literatur

Empfehlungen zum selben Thema automatisch vorgeschlagen von bX