The Validity and Absolute Reliability of Lower Extremity Angle Values on Full-Leg Standing Radiographs Using the TraumaMeter Software
Ist Teil von
  • Applied sciences, 2024-01, Vol.14 (2), p.565
Ort / Verlag
Basel: MDPI AG
Erscheinungsjahr
2024
Link zum Volltext
Quelle
EZB Free E-Journals
Beschreibungen/Notizen
  • To establish classifications and to obtain pre- and post-operative information on patient-specific alignments, it is necessary to measure different angular values accurately and precisely, mainly on weight-bearing, full-length anteroposterior X-rays of the lower limbs (LLRs). This study evaluated angular measurements’ validity and absolute reliability on LLRs with a self-developed, computer-aided measurement system (TraumaMeter v.873). Eight independent observers measured the preoperative mechanical hip-knee-ankle (mHKA) angle of 52 lower extremities (26 cases) in a blinded fashion on three occasions separated by two weeks. We obtained an intra-observer mean bias error (MBE) of 0.40°, a standard deviation (SD) of 0.11°, and a 95% confidence interval (CI) of 0.37°–0.43°. We also obtained an inter-observer MBE of 0.49°, an SD of 0.15°, and a 95% C of 0.45°–0.53°. The intra-observer MBE for the measurement pair between the second and the first measurement round (T2T1) was 0.43°, the SD was 0.13°, and the 95% CI was 0.39°–0.47°; the MBE between the third and the second round (T3T2) was 0.37°, with an SD of 0.10° and a 95% CI of 0.34°–0.40°; and the MBE between the third and the first round (T3T1) was 0.40°, with an SD of 0.10° and a 95% CI of 0.37°–0.43°. The interobserver MBE for the first round of measurements was 0.52°, with an SD of 0.16° and a 95% CI of 0.48°–0.56°; the MBE for the second round was 0.50°, with an SD of 0.15° and a 95% CI of 0.46°–0.54°; and the MBE for the third round was 0.46°, with an SD of 0.14° and a 95% CI of 0.42°–0.50°. There were no statistically significant differences in the inter-observer errors for the three tests. In the case of the intra-observer analysis, there were differences between T2T1 and between T3T2, but these differences were minimal, with no overlaps in the lower or upper values, respectively, of the confidence intervals. These results led us to conclude that the TraumaMeter v.873 software extension for measuring lower-limb angles in LLRs is an accurate tool with low intra- and inter-observer variability.
Sprache
Englisch
Identifikatoren
ISSN: 2076-3417
eISSN: 2076-3417
DOI: 10.3390/app14020565
Titel-ID: cdi_doaj_primary_oai_doaj_org_article_8fa5053b1e904f579eddf45fe3eb5288

Weiterführende Literatur

Empfehlungen zum selben Thema automatisch vorgeschlagen von bX