Compatibility of a novel temperature-controlled, irrigated radiofrequency catheter with ultra-high-density mapping
Ist Teil von
Journal of arrhythmia, 2023-08, Vol.39 (4), p.676-680
Ort / Verlag
Japan: John Wiley & Sons, Inc
Erscheinungsjahr
2023
Link zum Volltext
Quelle
Wiley Online Library Journals Frontfile Complete
Beschreibungen/Notizen
Compatibility of DiamondTemp (DT) radiofrequency (RF) catheter with the Rhythmia mapping system has not been manufacturer-reported nor its tracking accuracy reported.
Consecutive patients undergoing macroreentrant atrial tachycardia ablation guided by Rhythmia and ablated using DT were prospectively enrolled. Following catheter configuration, ablation lines were performed and remapped to measure the RF tag to effective-ablation-line-center (RFT-ALC) distance.
Among 20 consecutive patients (54 maps), 40 ablation lines were evaluated. Overall, the RFT-ALC distance was 3.88 ± 2.95 mm, and the operator assessment of accuracy was high. No complications occurred.
The use of DT catheter guided by the Rhythmia mapping system is feasible and accurate.