Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
We assessed feasibility, efficacy and safety of a suture-mediated closure device, Perclose Proglide (Abbott Vascular Devices, Santa Clara, CA, USA), for closure of the femoral vein access after percutaneous MitraClip (Abbott Vascular Devices) implantation.
Venous access of 80 consecutive patients undergoing percutaneous mitral valve repair using the MitraClip device was managed either by manual compression, "figure eight" suture and compression bandage for 12 hours, or by applying the Proglide device for haemostasis after the procedure (40 patients each group). Patients with Proglide closure showed complete immediate haemostasis in 92.5% (37/40) and were immobilised with a compression bandage for only four hours. In the Proglide group, one arteriovenous fistula was observed and had to be treated by vascular surgery. The overall duration of stay on an intensive care unit was significantly reduced in the Proglide group (59.4±48.9 hours vs. 84.6±59.5 hours, p<0.005).
Using a suture-mediated closure device for the femoral vein after percutaneous MitraClip implantation is feasible and safe. This allows earlier patient mobilisation and may reduce post-interventional duration of stay on an intensive care unit.