Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
Ergebnis 1 von 64

Details

Autor(en) / Beteiligte
Titel
An application of electrically cooled Si detector to fast neutral measurement on CHS
Ist Teil von
  • Review of Scientific Instruments, 2001-01, Vol.72 (1), p.615-618
Erscheinungsjahr
2001
Link zum Volltext
Quelle
Alma/SFX Local Collection
Beschreibungen/Notizen
  • A new type of fast neutral particle analyzer using Si diode cooled by Peltier module has been designed and developed. It is known that leak current in a solid-state detector is reduced by its use in low temperature. It is shown in this article that Peltier effect provides a very handy way of cooling the detector as long as a temperature around 0 °C is aimed at. The detector was easily cooled down to −1.6 °C and the performance of the detector was studied by the use of x/γ rays from 241 Am . The leak current was remarkably reduced as the detector is cooled improving energy resolution from 9.2 keV at room temperature to 4.3 keV at −1.6 °C. In addition, the lower detectable energy limit was expanded from 25 keV to 14 keV. A Si detector is also sensitive to visible light and x ray. Therefore, the detector of our “reference design” is furnished with an evaporated 1000-Å-thick aluminum layer to reduce the visible light and with a lead collimator to reduce the x ray. The detector was finally installed on the Compact Helical System device for a test of overall performance and an energy spectrum from a diagnostic neutral beam was successfully measured. This test was conducted without an Al layer and this result suggests a possibility of eliminating it.
Sprache
Englisch
Identifikatoren
ISSN: 0034-6748
eISSN: 1089-7623
DOI: 10.1063/1.1316761
Titel-ID: cdi_crossref_primary_10_1063_1_1316761
Format

Weiterführende Literatur

Empfehlungen zum selben Thema automatisch vorgeschlagen von bX