Effects of neutron energy spectrum on the efficiency calibration of epithermal neutron detectors
Ist Teil von
Review of scientific instruments, 1985-05, Vol.56 (5), p.1084-1086
Erscheinungsjahr
1985
Quelle
Alma/SFX Local Collection
Beschreibungen/Notizen
The 2
3
5U epithermal neutron fission detector system at TFTR has been calibrated i
n
s
i
t
u, using three different neutron sources: 2
5
2Cf, a D‐D, and a D‐T neutron generator. These sources produced a continuous spectrum, a line spectrum at ∼2.5 MeV, and a line spectrum at ∼14 MeV, respectively. The toroidally averaged detection efficiencies are 4.61×10−
9 (2
5
2Cf), 4.80×10−
9 (D‐D), and 5.42×10−
9 (D‐T) counts/neutron for a moderated 1.3‐g 2
3
5U detector placed ∼4 m from the torus centerline. The effects of (n,2n) and (n,n
’γ) reactions are discussed.