European Journal of Nuclear Medicine, 1989, Vol.15 (11), p.732-735
1989
Volltextzugriff (PDF)

Details

Autor(en) / Beteiligte
Titel
Study on the stability of the calibration and normalization in PET and the influence of drifts on the accuracy of quantification
Ist Teil von
  • European Journal of Nuclear Medicine, 1989, Vol.15 (11), p.732-735
Ort / Verlag
Germany
Erscheinungsjahr
1989
Quelle
springer (창간호~2014)
Beschreibungen/Notizen
  • With regard to the quality control of quantification in positron emission tomography some characteristics were examined, to develop a simple method for frequent monitoring. The stability and uniformity of the detector count rate was checked by plotting the RMS deviation of the non normalized count rate and the standard deviation of the normalized count rate, each normalized to its value after the calibration or normalization respectively. Switching off a single detector did not impair the image quality, but the normalized image pixel counts were reduced by 2%-3% when a detector block was switched off. Thus in case of need, a weak detector can still be used to perform a scan. A reduced count rate capability at specific activities above 5 x 10(4) Bq/ml (approximately 1.5 x 10(5) corrected true system counts/s or approximately 1.5 x 10(4)/s for a plane) was found compared to the maximum usable activity of 8 x 10(4) Bq/ml obtained 1 year earlier, indicating a drift in the count loss corrections. A variation of the room temperature changes the temperature distribution inside the gantry by 5% per degree C and the drift of the sensitivity (normalized image pixel counts) is 2% per degree C.
Sprache
Englisch
Identifikatoren
ISSN: 0340-6997
eISSN: 1619-7089
DOI: 10.1007/bf00631766
Titel-ID: cdi_crossref_primary_10_1007_BF00631766

Weiterführende Literatur

Empfehlungen zum selben Thema automatisch vorgeschlagen von bX