Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
Collimator design for single photon emission tomography
Ist Teil von
European Journal of Nuclear Medicine, 1992, Vol.19 (2), p.138
Ort / Verlag
Germany
Erscheinungsjahr
1992
Quelle
MEDLINE
Beschreibungen/Notizen
We discuss recent trends in collimator design and technology, with emphasis on theoretical and practical issues of importance for single photon emission tomography (SPET). The well-known imaging performance parameters of parallel-hole collimators are compared with those of fan-beam collimators, which have enjoyed considerable success in recent years, particularly for brain SPET. We review a simplistic approach to the collimator optimization problem, as well as more sophisticated "task-dependent" treatments and important considerations for SPET collimator design. Practical guidance is offered for understanding trade-offs that must be considered for clinical imaging. Finally, selective comparisons among different SPET systems and collimators are presented for illustrative purposes.