Sie befinden Sich nicht im Netzwerk der Universität Paderborn. Der Zugriff auf elektronische Ressourcen ist gegebenenfalls nur via VPN oder Shibboleth (DFN-AAI) möglich. mehr Informationen...
Simultaneous high-resolution cardiac T 1 mapping and cine imaging using model-based iterative image reconstruction
Ist Teil von
Magnetic resonance in medicine, 2019-02, Vol.81 (2), p.1080-1091
Ort / Verlag
United States
Erscheinungsjahr
2019
Link zum Volltext
Quelle
MEDLINE
Beschreibungen/Notizen
To provide high-resolution cardiac T
mapping of various cardiac phases and cine imaging within a single breath-hold using continuous golden ratio-based radial acquisition and model-based iterative image reconstruction.
Data acquisition was performed continuously using golden ratio-based radial sampling and multiple inversion pulses were applied independent of the heart rate. Native T
maps of diastole and systole were reconstructed with in-plane resolution of 1.3 × 1.3 mm
using model-based iterative image reconstruction. Cine images with 30 cardiac phases were reconstructed from the same data using kt-SENSE. The method was evaluated in a commercially available T
phantom and 10 healthy subjects. In vivo T
assessment was carried out segment-wise.
Evaluation in the phantom demonstrated accurate T
times (R
> 0.99) and insensitivity to the heart rate. In vivo T
values did not differ between systole and diastole, and T
times assessed by the proposed approach were longer than measured with a modified Look-Locker inversion recovery (MOLLI) sequence, except for lateral segments. Cine images had a consistent dark-blood contrast and functional assessment was in agreement with assessment based on Cartesian cine scans (difference in ejection fraction: 0.26 ± 2.65%, P = 0.65).
The proposed approach provides native T
maps of diastole and systole with high spatial resolution and cine images simultaneously within 16 s, which could strongly improve the scan efficiency.